9
2.
INSTALLATION
2.1
Unpacking
The Lambda HPX and associated hardware comes packed in a single carton. The following is
a list of the components found there. If you believe that any of these components are missing
or show obvious signs of damage from shipping, please contact the factory.
1.
Controller cabinet with factory-installed high-power LED assembly, control circuitry, and
power supply.
2.
Power Cord
3.
Manual
4.
Adapter
5.
One or more additional High-Power LED Assemblies containing different wavelength
LEDs, if ordered.
6.
Small Zeiss female dovetail
7.
Liquid Light Guide dovetail with C-mount extension
8.
Liquid Light Guide (if ordered)
9.
Microscope adapters (if ordered)
2.2
Installation Instructions
1.
Turn power switch to OFF (“0”) position.
2.
Plug power cable into the mains outlet.
3.
In the typical application, the collimated light output of the Lambda HPX-L5 is focused
onto the input end of a liquid light guide. The lamp’s integral cold mirror allows the beam
to be focused without danger to the light guide.
The liquid light guide has a black “end sleeve” preinstalled over the brass ferrule on each
end. The two ends of the light guide are identical. The end sleeves are designed to slide
into the mounting flanges on the input and output lens tubes.
The lens tube on the input end of the light guide comes with the guide and looks like that
shown in the figure. The end sleeve slides into one end of the tube while the other end
holds a UV-grade, fused-silica lens system (25.4 mm diameter, 33 mm focal length). The
position of the sleeve in the tube can be later adjusted to optimize the light output from
the light guide.
The lens tube slides into the dovetail tube. The dovetail tube is used to fasten the light
guide and lens tube to the mating female adapter mounted either directly at the output of
the lamp or at the output of a filter wheel (location and control external to the Lambda
HPX-L5).
The output lens system, at the opposite end of the light guide determines the
characteristics of the output light. The output lens, lens tube, and mounting hardware
are integral parts of the microscope adapter system when ordered for the liquid light
guide. In many cases, the output parts are nearly identical to the input parts.
LAMBDA HPX-L5 OPERATION MANUAL – REV. 1.03 (20161130)
Содержание Lambda HPX-L5
Страница 3: ......
Страница 4: ......
Страница 16: ...LAMBDA HPX L5 OPERATION MANUAL REV 1 03 20161130 14 This page intentionally blank...
Страница 20: ...LAMBDA HPX L5 OPERATION MANUAL REV 1 03 20161130 18 This page intentionally blank...
Страница 22: ...LAMBDA HPX L5 OPERATION MANUAL REV 1 03 20161130 20 This page intentionally blank...
Страница 24: ...LAMBDA HPX L5 OPERATION MANUAL REV 1 03 20161130 22 This page intentionally blank...
Страница 26: ...LAMBDA HPX L5 OPERATION MANUAL REV 1 03 20161130 24 This page intentionally blank...
Страница 28: ...LAMBDA HPX L5 OPERATION MANUAL REV 1 03 20161130 26 This page intentionally blank...