13
2.3.4
Firmware and Lambda HPX Model (Switch 4)
The position of DIP Switch 4 informs the control board’s firmware of the specific Lambda
HPX model it is controlling. This witch must be ON (Lambda HPX-L5 ( air-cooled lamp with
liquid light guide)).
Table 2-1. DIP Switch configuration settings.
Switch Definition Position
Definition
ON*
Same as TTL IN:
TTL IN=low
Æ
TTL OUT=low;
TTL IN=high
Æ
TTL OUT=high
1
TTL OUT logic
OFF
Inverse of TTL IN:
TTL IN=low
Æ
TTL OUT=high;
TTL IN=high
Æ
TTL OUT=low
ON
TTL IN=high turns lamp on
2
TTL IN logic
OFF*
TTL IN=low turns lamp on
ON*
Lamp turns on (Notes 1 and 2)
3
Lamp state when switching
from manual control to TTL
OFF
Lamp turns off (Note 1)
ON*
Lambda HPX-L5
4
Lambda HPX model-specific
operation (Note 3)
OFF Lambda
HPX
“*” = Default
NOTE 1: After switching from manual mode (LED switch position ON or OFF) to TTL mode
(LED switch position TTL), TTL control of the lamp’s on/off state will resume with the next
change of logic state at TTL IN.
NOTE 2: While using the LED switch to switch from ON to TTL, the lamp turns off while
transitioning through the OFF position (the OFF position is located in between ON and
TTL).
NOTE 3: Switch 4 must be ON for proper operation of the Lambda HPX-L5 (forced air-
cooled lamp with liquid light guide).
LAMBDA HPX-L5 OPERATION MANUAL – REV. 1.03 (20161130)
Содержание Lambda HPX-L5
Страница 3: ......
Страница 4: ......
Страница 16: ...LAMBDA HPX L5 OPERATION MANUAL REV 1 03 20161130 14 This page intentionally blank...
Страница 20: ...LAMBDA HPX L5 OPERATION MANUAL REV 1 03 20161130 18 This page intentionally blank...
Страница 22: ...LAMBDA HPX L5 OPERATION MANUAL REV 1 03 20161130 20 This page intentionally blank...
Страница 24: ...LAMBDA HPX L5 OPERATION MANUAL REV 1 03 20161130 22 This page intentionally blank...
Страница 26: ...LAMBDA HPX L5 OPERATION MANUAL REV 1 03 20161130 24 This page intentionally blank...
Страница 28: ...LAMBDA HPX L5 OPERATION MANUAL REV 1 03 20161130 26 This page intentionally blank...