6.2.4 Communication Parameters (SErP)
No.
Sign
Range
Default
Description
Set
800
Ad
0~99
1
Communication Address
801
802
b1
b2
0~2
0
0
COM1/COM2 Baud Rate
0
:
9600bps
1
:
19200bps
2: 115200bps [1152d]
803
804
P1
P2
0~2
0
0
COM1/COM2 Parity Check
0. none
[
None Check]
1. EVEn
[Even Check]
2. odd
[Odd Check]
805
806
E1
E2
0~2
0
0
COM1/COM2 Communication Mode
0. HASC
[Host-slave, Modbus ASCII]
1. Hrtu
[Host-slave, Modbus RTU]
2. Cont
[Continuous Sending ASCII]
807
dA
0~3
2
Data for Continuous Sending Mode
0. groS
[Gross Weight]
1. nEt
[Net Weight]
2. dISP
[Displayed Characters]
3. PnEt
[Net Weight Peak Value]
808
SF
0~9
2
Continuous Sending Frequency
[Hz]
0: 1Hz; 1: 2Hz; 2: 5Hz; 3: 10Hz
4: 20Hz; 5: 25Hz; 6: 50Hz; 7: 100Hz
8: 500Hz; 9: 1000Hz
COM1 Baud Rate
Max. Sending Frequency
9600bps
50Hz
19200bps
100Hz
115200bps
1000Hz
809
dF
0~3
0
Modbus Data Format
Reading&Writing Order of 4-Byte Registers:
0: 4321
[HB4 HB3 LB2 LB1]
1: 3412
[HB3 HB4 LB1 LB2]
2: 1234
[LB1 LB2 HB3 HB4]
3: 2143
[LB2 LB1 HB4 HB3]
The HEX byte order of float and long int registers
in the controller is
‘
HB4 HB3 LB2 LB1
’
.