6.2.2 Calibration Parameters (CALP)
No.
Sign
Range
Default
Description
Set
120
ZP
0~1
0
Auto Zero Tracking Permission
0: oFF
1: on
(Only when weight is stable, Auto Zero
Tracking will be allowed. Refer to Parameter
[106]/[107])
121
Zt
1~10
1
Auto Zero Tracking Time T
[s]
T = Set value × [107] Stablity Judging Time.
122
Zr
0.1~50.0
0.2
Auto Zero Tracking Range
[Division]
Zero Tracking Rate = [122] / [121].
Suggestion: Zero Tracking Rate
≤
0.5[Division/s].
123
nZ
0~50000
50
Zero Fine Adjusting Range
If Gross Weight variation caused by Zero Value
changing is within this range,
‘Zero Fine
Adjustment’
will be valid.
Suggestion: Set value
≤
(Scale Capacity × 4%).
124
dZ
0~99
5
Range of Displaying Zero
When the absolute value of weight is within this
range, the weight display value will return to zero.
125
tL
1~99999
12000
[*]
Total Capacity of Loadcells
Total Capacity of Loadcells = Loadcell Capacity ×
Loadcell Number.
Only for query.
126
SE
0.500~5.000
2.000
[*]
Output Sensitivity of Loadcell
[mV/V]
Only for query.
127
Un
0
0
Unused
128
In
0~1
0
Initial Zero Adjustment after Power-on
0: oFF
1: on
[without Power-down Protection]
129
It
0~1800
10
Initial Zero Adjusting Time
[s]
130
Ir
0~5000
50
Initial Zero Adjusting Range
Suggestion: Set value
≤
(Scale Capacity × 20%).
[*]:
‘
RAM Reset
’
operation has no effect on the parameter.