42
TROUBLESHOOTING
TROUBLESHOOTING
TROUBLESHOOTING
TROUBLESHOOTING
The troubleshooting chart provides pointers on diagnosing issues associated with error or
status codes.
Problem: The
ADvew
will not power on.
Solutions:
1.
The
ADview
must be plugged in to charge before first use.
Problem:
My
ADview
is not communicating with my EMR/EHR system or network
Solutions:
1.
Be sure your
ADview
is equipped with the Bluetooth option.
2.
Your
ADview
and wireless system must be configured to communicate.
Refer to your EMR/EHR operator’s manual or contact your IT administrator
or EMR/EHR provider.
3.
You may need to move your ADview closer to the wireless system. While
the Bluetooth capability on the ADview has been rated up to a maximum
range of 100 m, be sure to minimize walls, structures and other obstacles
that may impede wireless connectivity.
Troubleshooting - Blood Pressure Module
Problem: Wrong size cuff, Misplaced cuff, or Blocked brachial artery
Solutions:
1.
Check that the cuff is in the correct position.
2.
Check that the cuff is properly tightened.
3.
Check that there is no excessive clothing between the arm and the cuff.
4.
Check that the cuff applied is of the correct size.
5.
The patient may have been moving too much.
6.
Take another BP reading.
Problem: Too much patient or environment motion or conditions causing tremors
Solutions:
1.
Check that the cuff is in the correct position.
2.
The patient may have been moving too much.
3.
Take another BP reading.