32
MANAGING READINGS
MANAGING READINGS
MANAGING READINGS
MANAGING READINGS
Recalling the Last
Recalling the Last
Recalling the Last
Recalling the Last Set of Readings
Set of Readings
Set of Readings
Set of Readings
To redisplay the last set of readings, depress the automatic button
on the BP module for
more than two seconds until the last reading is displayed. If your
ADview
has temperature
and/or pulse oximetry modules, the last set of readings includes these readings as well. If the
last attempted reading resulted in an error and/or warning, then this will be displayed. The
device will display dashes if no reading is in memory, a reading was aborted or, the previous
BP was a Sphygmode measurement.
For the pulse oximetry module: In the event of the 10-minute measurement timeout, the
module will terminate the measurement and status code “01” will display on the main
module. The last valid reading recorded at the end of the ten-minute period will be the
recalled reading.
Clearing the Last Set of Readings
Clearing the Last Set of Readings
Clearing the Last Set of Readings
Clearing the Last Set of Readings
To clear the values from the last automatic BP measurement and the accessories, press and
hold the automatic button
more than 5 seconds. Previous values will be displayed
momentarily. Then the display blanks. On redisplay you will see dashes for all the values
that have been cleared.
Your
ADview
displays the most recent set of readings for two minutes. If
patient privacy is a concern, you can clear these readings from the display
before collecting vital signs from another patient.
Bluetooth Wireless Transmission of Readings
Bluetooth Wireless Transmission of Readings
Bluetooth Wireless Transmission of Readings
Bluetooth Wireless Transmission of Readings
If your
ADview
has the optional Bluetooth wireless communication capability you may be
able to transmit the readings into your EMR/EHR system or communication network.
Contact your IT administrator or EMR/EHR vendor for support on configuring your
ADview
for Bluetooth wireless communication.