35
3.
Compare the pressure reference to the
ADview
throughout the pressure range, 0 to
270mmHg. If the difference between the pressure reference and the
ADview
is no
larger than 2mmHg, the
ADview
is calibrated correctly for operation. If the
ADview
needs calibrating, contact an authorized service center on page 48.
4.
Exit the calibration check by pressing the automatic button
again. Once the
display shows dashes as the systolic BP, you are now ready to take a measurement.
It is recommended that you check the temperature calibration of your
ADview
temperature
module annually. A Calibration Plug (part# 9000TCP) is available to check accuracy of the
technology. It replaces the regular probe and verifies the accuracy of the temperature
electronics. Replace the temperature module if error is greater than +/-0.1° C. To check the
calibration:
1.
Remove the Temperature Well from the Temperature Module.
2.
Replace the Temperature Probe with the Calibration Plug.
3.
Replace the Temperature Well.
4.
Initiate a measurement by inserting and removing the Temperature probe from the
well.
5.
Verify the accuracy of the measurement.
Probe accuracy can be checked by using the thermometer as you would on a patient in any
mode with a cover, but place it in a calibrated water bath. Direct mode is accurate to +/- 0.1°
C of the calibrated water bath temperature. Predictive mode accuracy is accurate to +/- 0.2° C
of the calibrated water bath temperature.