34
3.
Dry sensor and clips thoroughly before re-use.
4.
Never immerse sensor and clips in fluids. Do not pour or spray any liquids
on them either. Caustic or abrasive cleaners will cause permanent damage.
Prying the finger clip sensor to an angle greater than 90° can permanently
damage its casing.
Checking the Calibration of Your Device
It is recommended that you check the BP calibration of your
ADview
once a year. To check
calibration:
1.
Start with the device powered off. While holding down the automatic button
on the front of the BP module, toggle on the power button on the right side of the
main enclosure. The “CAL” message is displayed in the pulse rate display to indicate
that the system is in the calibration mode. During this mode, the system pressure
displays in the systolic BP display area.
2.
Using a set of two T-connectors, connect a calibrated pressure reference and control,
such as a manometer of known accuracy, and inflation bulb, to the pressure hose
connector of the
ADview
. Contact ADC customer support for details on ordering the
calibration kit which includes a T-connector. The diagram below shows how these
four components should be connected.