MATTRESS CLEANING INSTRUCTIONS:
CAUTION:
A.
It is recommended that gloves and protective clothing be worn at all times during
cleaning and disinfecting
B.
Hypercarbonate and phenol based solutions
should not
be used.
During Operation:
For normal soil, wipe down cover and interior sacs with a standard cleaning
disinfectant and wipe off excess moisture with a towel.
Laundering Mattress:
1.
Disconnect the power and deflate the interior sacs.
2.
Disassemble cover, interior sacs, connecting hoses and base. Place them in
different laundering baskets.
3.
Spray disinfectants and wipe down interior sacs. Air-dry sacs.
DO NOT expose
to heat.
4.
Spray disinfectants and wipe down foam cover.
5.
Place cover and base in washing machine.
6.
Use standard hospital disinfectant/detergent.
7.
Set washing cycles as recommended below:
Regular Cycle
a)
Pre-soak with disinfectant/detergent with diluted bleach in cold water (10
minutes.)
b)
Main wash for 15 minutes (Time depending on soil level).
c)
Rinse Cycle: 5 minutes minimum.
d)
Spin/Drain Cycle: 5 minutes minimum.
e)
Air-dry or place in dryer at very low or no heat.
8.
After the cover, interior sacs, and base are dry, store them in a sterilized bag and
seal for next use.
Notice:
DO NOT use water temperatures greater than 150°F (65°C).
BLOWER CLEANING INSTUCTIONS:
CAUTION:
A.
It is recommended that gloves and protective clothing be worn at all times during
cleaning and disinfecting
B.
Hypercarbonate and phenol based solutions
should not
be used.
C.
Make sure unit is turned “off” and unplugged before cleaning.
For normal soil, wipe down with a standard cleaning spray and wipe off excess moisture
with a towel. Spray with a disinfectant and let it stand for a few minutes before wiping down.
Check corners and creases for soil build-up. Look to see if any external parts are damaged or
need replacing such as: hooks, foot pads and filter. Replacement parts can be purchased. See last
page for parts listing.