GENERAL WARNINGS AND
PRECAUTIONS
WARNINGS
WARNING:
DO NOT use this device if the power cord is cut, frayed or loosely
connected to the device.
WARNING:
Make sure power cord is plugged into a properly grounded A/C 110V
outlet.
PRECAUTIONS
CAUTION:
DO NOT use around an open flame.
CAUTION:
The housing of the master control unit should only be opened by qualified
technical personnel. Before opening the housing, make sure the unit is turned off and
unplugged from its power source.
CAUTION:
Electric Powered Mechanism. Electrical Hazard may occur if device is
plugged into inadequate electrical outlet. To avoid electrical shock hazard, make sure
unit is plugged into a grounded A/C 110 Volt outlet.
PURPOSE OF THE DEVICE
The purpose of the Sapphire Series 1100 & 1100EC Mattress System is to provide therapeutic
benefit to patients at risk or suffering from pressure ulcers.
The active component that has contact with the patient is a specialized, multi-cell air mattress
sized to fit a standard medical bed frame. The air mattress serves to replace the original mattress
and is equipped with 4 air hoses with connectors that mate with the Air Blower/Control Unit.
The Control Unit is a self contained, totally enclosed module that hangs by retractable hooks on
the bed frame at the foot of the bed or sits on the floor under the bed.
It is provided with a hospital grade electrical cord and a control panel that has selector switches
and indicator lights. The switches and indicators are protected under a flexible membrane to
keep out liquid spills and enhance clean-up and sanitation. Inside the Control Unit is a variable
output blower, a valve motor and air valves that allow the air mattress to operate in static mode
or provide alternating pressure variations within the mattress. There is also a printed circuit
board, which provides the electrical controls. The Control Unit has one user replaceable 250v
5.0 amp fuse near the AC Power Cord Entry.
3