![Sumitomo HC-4E1 Скачать руководство пользователя страница 6](http://html1.mh-extra.com/html/sumitomo/hc-4e1/hc-4e1_technical-manual_1390618006.webp)
Safety
2
WARNINGS
AVOID ELECTRIC SHOCK.
All electrical supply equipment must meet applicable codes and
be installed by qualified personnel. Permit only qualified electrical technicians to open electrical
enclosures, to perform electrical checks or to perform tests with the power supply connected
and wiring exposed. Failure to observe this warning can result in serious injury or death.
Disconnect the power to the compressor before troubleshooting the electrical components.
AVOID INJURY.
Never use compressed helium gas from a cylinder without a proper regulator.
Overpressure can cause serious injury if the system equipment ruptures.
During operation, some surfaces under the compressor’s cover become hot. Allow the
compressor to cool for 1/2 hour after shutdown before removing the cover for maintenance.
Always wear eye protection when handling pressurized gas lines and other pressurized
equipment. Never apply heat to a pressurized gas line or other pressurized components.
Disconnect gas lines only when the compressor is stopped. Disconnecting the cold head while
it is cold can create excessively high internal pressure as the gas warms. Material failure and
uncontrolled pressure release can cause injury.
Use two wrenches when disconnecting a gas line coupling to avoid loosening the cold head or
compressor coupling. Gas pressure can project the coupling with enough force to cause
serious injury.
The compressor is charged with helium gas. Vent both supply and return Aeroquip couplings to
atmospheric pressure before disassembly, except when disconnecting the adsorber or the gas
lines. Uncontrolled pressure release can cause serious injury.
Always vent a gas-charged component before beginning to disassemble its couplings. Gas
pressure can launch a loose coupling with enough force to cause serious injury.
The adsorber is charged with helium gas. Follow the used adsorber venting procedure for safe
disposal of the used adsorber.
CAUTIONS
PRESERVE YOUR WARRANTY.
Modification to equipment without the consent of the
manufacturer will void the warranty.
Specifications require the use of 99.995% pure helium gas. Using a lesser quality of helium can
damage the system and void the warranty.
AVOID GAS LEAKS.
Check the condition of the gasket face seal on the male half of each
Aeroquip coupling. Be sure the gasket face seal is in place and the sealing surfaces on both
the male and female halves are clean before connecting. Replace the gasket face seal if it is
damaged or missing.
Keep the gas line couplings aligned when making or breaking a coupling connection. Leaks can
occur due to the weight of the gas line or due to a sharp bend near the connection.
Содержание HC-4E1
Страница 2: ......
Страница 8: ...4 This page is intentionally blank...
Страница 10: ...6 This page is intentionally blank...
Страница 12: ...8 This page is intentionally blank...
Страница 20: ...16 This page is intentionally blank...
Страница 26: ...22 This page is intentionally blank...
Страница 38: ...34 This page is intentionally blank...
Страница 55: ...Troubleshooting 51 Figure 9 HC 4E1 Wiring Diagram...
Страница 56: ...Troubleshooting 52 Figure 10 HC 4E2 Wiring Diagram...
Страница 57: ...Troubleshooting 53 Figure 11 HC 4E1 Electrical Schematic...
Страница 58: ...Troubleshooting 54 Figure 12 HC 4E2 Electrical Schematic...
Страница 61: ...Parts 57 Figure 13 Parts Identification 18 21 19 13 14 20 16...