Troubleshooting
48
Problem
Possible Cause
Corrective Action
Gradual loss of helium gas
pressure.
Gas is leaking from the
compressor.
Leak check the compressor and
repair.
Compressor runs; elapsed
time meter does not run.
Defective elapsed time meter
or motor contactor.
Consult a SCAI Service Center.
Fan runs slow or has stopped Open fuse in fan/circulator
circuit
Check for open fuse. Replace if
open.
Defective transformer in
fan/circulator circuit
Check winding resistance and
for ground fault. If defective,
replace transformer
Defective fan motor
Check winding resistance and
for ground fault. If defective,
replace fan.
Check for free rotation when
power is off. If rotation is not
free, replace fan.
Shorted run capacitor
Check capacitor resistance. If
defective, replace capacitor.
Open internal motor protector
Let fan cool and restart
Fan blade obstruction
Remove obstruction
Coolant flow has stopped
(Coolant piping between
compressor motor and air
heat exchanger does not get
warm when compressor is
running).
Open fuse in fan/circulator
circuit
Check for open fuse. Replace if
open.
Defective transformer in
fan/circulator circuit
Check winding resistance and
for ground fault. If defective,
replace transformer
Defective circulator motor
Check winding resistance and
for ground fault. If defective,
consult a SCAI Service Center.
Shorted run capacitor
Check capacitor resistance. If
defective, replace capacitor.
Circulator bearing failure
Consult a SCAI Service Center.
Содержание HC-4A
Страница 2: ......
Страница 6: ...iv This page is intentionally blank ...
Страница 10: ...4 This page is intentionally blank ...
Страница 12: ...6 This page is intentionally blank ...
Страница 14: ...8 This page is intentionally blank ...
Страница 36: ...30 This page is intentionally blank ...
Страница 38: ...32 This page is intentionally blank ...
Страница 59: ...Troubleshooting 53 Figure 8 HC 4A Wiring Diagram Main Electrical Chassis and Components ...
Страница 60: ...Troubleshooting 54 Figure 9 HC 4A2 Wiring Diagram Main Electrical Chassis and Components ...
Страница 61: ...Troubleshooting 55 Figure 10 HC 4A and HC 4A2 Wiring Diagram Air Cooled Unit ...
Страница 62: ...Troubleshooting 56 Figure 11 HC 4A Electrical Schematic ...
Страница 63: ...Troubleshooting 57 Figure 12 HC 4A2 Electrical Schematic Main Electrical Chassis and Components ...
Страница 64: ...Troubleshooting 58 Figure 13 HC 4A2 Electrical Schematic Air Cooled Unit ...
Страница 67: ...Parts 61 Figure 14 Parts Identification Fan Circulator and Air Heat Exchanger not shown 15 18 16 10 11 17 13 ...
Страница 68: ...Parts 62 Figure 15 Parts Identification for HC 4A 12 14 19 23 ...
Страница 69: ...Parts 63 Figure 16 Parts Identification for HC 4A2 12 14 19 23 ...
Страница 70: ...Parts 64 Figure 17 Parts Identification 29 28 30 31 ...