MAINTENANCE
33
AVOID INJURY.
Disconnect gas lines only when the compressor is stopped.
Disconnecting the cold head while it is cold may create excessively high internal
pressure as the gas warms. Material failure and uncontrolled pressure release
can cause serious injury.
AVOID INJURY.
The compressor is charged with helium gas. Vent the com -
pressor to atmospheric pressure before disassembly, except when disconnecting
adsorber or the gas lines. Uncontrolled pressure release can cause serious
injury.
AVOID INJURY.
Never use compressed helium gas from a cylinder without a
proper regulator. Overpressure can cause serious injury if the system equipment
ruptures.
AVOID INJURY FROM BURNS.
During operation, some surfaces in the
compressor capsule become hot. Allow the compressor to cool for 1/2 hour after
shutdown before removing the cover for maintenance.
AVOID INJURY.
Rotating fan blade can cause injury. Disconnect power before
removing compressor covers for maintenance.
AVOID INJURY.
When handling pressurized gas lines and other pressurized
equipment, always wear eye protection. Never apply heat to a pressurized gas
line or other pressurized components.
AVOID INJURY.
Coolant fluid is a 30% ethylene glycol/water mixture.
TOXIC: Do not ingest. Contact local environmental authorities for proper
disposal of ethylene glycol mixtures.
PRESERVE YOUR WARRANTY.
Modification to equipment without the consent
of the manufacturer will void the warranty.
Содержание HC-4A
Страница 2: ......
Страница 6: ...iv This page is intentionally blank ...
Страница 10: ...4 This page is intentionally blank ...
Страница 12: ...6 This page is intentionally blank ...
Страница 14: ...8 This page is intentionally blank ...
Страница 36: ...30 This page is intentionally blank ...
Страница 38: ...32 This page is intentionally blank ...
Страница 59: ...Troubleshooting 53 Figure 8 HC 4A Wiring Diagram Main Electrical Chassis and Components ...
Страница 60: ...Troubleshooting 54 Figure 9 HC 4A2 Wiring Diagram Main Electrical Chassis and Components ...
Страница 61: ...Troubleshooting 55 Figure 10 HC 4A and HC 4A2 Wiring Diagram Air Cooled Unit ...
Страница 62: ...Troubleshooting 56 Figure 11 HC 4A Electrical Schematic ...
Страница 63: ...Troubleshooting 57 Figure 12 HC 4A2 Electrical Schematic Main Electrical Chassis and Components ...
Страница 64: ...Troubleshooting 58 Figure 13 HC 4A2 Electrical Schematic Air Cooled Unit ...
Страница 67: ...Parts 61 Figure 14 Parts Identification Fan Circulator and Air Heat Exchanger not shown 15 18 16 10 11 17 13 ...
Страница 68: ...Parts 62 Figure 15 Parts Identification for HC 4A 12 14 19 23 ...
Страница 69: ...Parts 63 Figure 16 Parts Identification for HC 4A2 12 14 19 23 ...
Страница 70: ...Parts 64 Figure 17 Parts Identification 29 28 30 31 ...