![Sumitomo HC-4A Скачать руководство пользователя страница 49](http://html1.mh-extra.com/html/sumitomo/hc-4a/hc-4a_technical-manual_1390617049.webp)
Maintenance
43
Figure 6 Compressor Shell High Temperature Switch
Replace Fuses
Tools required: ¼” Flat blade screwdriver
#2 Philips screwdriver
Fuses for the HC-4A and HC-4A2 are located in three places:
1.
Front Panel
a.
Auxiliary power, solenoid valve, coolant circulator, fan circuit: (2) 5.0 ampere
b.
Cold head valve motor circuit: (2) 0.63 ampere [HC-4A] or (2) 1.0 ampere [HC-4A2]
2.
Behind left side compressor cover:
a.
Coolant circulator and fan circuit: (2) 1.0 ampere and (1) 0.63 ampere
3.
Inside electrical enclosure:
a.
Motor control circuit: (2) 0.4 ampere
AVOID ELECTRIC SHOCK.
Disconnect the power to the compressor before
replacing fuses.
AVOID ELECTRIC SHOCK.
Permit only qualified electrical technicians to open
electrical enclosures, to perform electrical checks or to perform tests with the
power supply connected and wiring exposed. Failure to observe this warning can
result in serious injury or death.
Front Panel:
1.
Open fuse holder by grasping cap with fingers, push and rotate counterclockwise.
Terminal Box
Cover
Temperature
Overload
Switch in
Retainer
Содержание HC-4A
Страница 2: ......
Страница 6: ...iv This page is intentionally blank ...
Страница 10: ...4 This page is intentionally blank ...
Страница 12: ...6 This page is intentionally blank ...
Страница 14: ...8 This page is intentionally blank ...
Страница 36: ...30 This page is intentionally blank ...
Страница 38: ...32 This page is intentionally blank ...
Страница 59: ...Troubleshooting 53 Figure 8 HC 4A Wiring Diagram Main Electrical Chassis and Components ...
Страница 60: ...Troubleshooting 54 Figure 9 HC 4A2 Wiring Diagram Main Electrical Chassis and Components ...
Страница 61: ...Troubleshooting 55 Figure 10 HC 4A and HC 4A2 Wiring Diagram Air Cooled Unit ...
Страница 62: ...Troubleshooting 56 Figure 11 HC 4A Electrical Schematic ...
Страница 63: ...Troubleshooting 57 Figure 12 HC 4A2 Electrical Schematic Main Electrical Chassis and Components ...
Страница 64: ...Troubleshooting 58 Figure 13 HC 4A2 Electrical Schematic Air Cooled Unit ...
Страница 67: ...Parts 61 Figure 14 Parts Identification Fan Circulator and Air Heat Exchanger not shown 15 18 16 10 11 17 13 ...
Страница 68: ...Parts 62 Figure 15 Parts Identification for HC 4A 12 14 19 23 ...
Страница 69: ...Parts 63 Figure 16 Parts Identification for HC 4A2 12 14 19 23 ...
Страница 70: ...Parts 64 Figure 17 Parts Identification 29 28 30 31 ...