Chapter 5 Specifications
The test for specifications should be carried out under the working temperature, warm up at least
30 minutes after power on, and make sure device is within calibration period. Please make
instrument placed separately and no stacked, good ventilation cooling and keep at least 150mm
airspace on both sides and rear side.
5.1 Continuous Output
5.1.1 Built-in Waveform:
165 kind of waveforms including Sine, Square, Ramp, Pulse, Noise, DC, PRBS
(Pseudorandom
Binary Sequence), User-defined Arbitrary, User-defined Harmonic and so on.
Sample rate: 250MSa/s
Amplitude resolution: 16 bits
5.1.2 Sine
Harmonic Distortion (0dBm):
≤-
60dBc frequency
≤
2
0MHz
≤-
50dBc frequency
≤
50MHz
≤-
40dBc frequency>50MHz
Total Distortion:
≤
0.075%(10Hz~20kHz
,
20Vpp)
5.1.3 Square
Duty Cycle:
0.0001%~99.9999%
(limited by frequency setting)
Overshoot:
≤
2%(1MHz, 1Vpp, typical value)
Edge time:
≤
4.2ns
5.1.4 Pulse
Duty Cycle:
0.0001%~99.9999%
(limited by settings of frequency and edge time)
Overshoot:
≤
2%(1MHz, 1Vpp,typical value)
Edge time:
≥
8ns, the rising edge and the falling edge are separately adjustable
Pulse width: 13ns ~ (period - 13ns) (limited by settings of frequency and edge time)