101515 REV C
1-21
Specifications
Classification
Class I equipment with Type BF applied part suitable for continuous operation. Not
classified for protection against ingress of liquid. Not classified for use in the presence
of flammable anesthetics.
Size (approx.)
11.5” x 8” x 8”
29.2cm x 20.3cm x 20.3cm
Weight
6.5 lb (2.9 kg) when empty
9 lb (4.0 kg) with unit filled with water to heating level
Reservoir capacity
93 oz (2.75 l) maximum
Flow rate
7 gph (26.5 lph) minimum with pad attached
Ambient operating temperature
60°F to 90°F (15.6°C to 32.2°C )
Environmental conditions for transport
and storage
-20°F to 120°F (-28°C to 48°C )
At uncontrolled RH
Temperature setpoints
TP650 Classic
107°F (42°C) High heat
100°F (38°C) - Low heat
50°F (10°C) - Cooling
TP700 Professional
107°F (42°C)
100°F (38°C)
95°F (35°C)
50°F (10°C)
Average temperature accuracy
±2°F at 107°F (±1°C at 42°C)
Maximum Contact Surface
Temperature
107°F (42°C)
High Limit Safety Temperature
110°F to 120°F
(43.3°C to 49°C)
Power cord
International (harmonized) 3-wire cordset
Current leakage
100 microamperes maximum
Ground resistance
0.5 ohm max
Electrical requirements
Voltage (VAC)
Frequency (Hz)
Current (amps)
120±10%
60
3.1 amperes
Certifications
MEDICAL ELECTRICAL EQUIPMENT WITH RESPECT TO ELECTRICAL SHOCK, FIRE,
AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 60601-1, AND CAN/CSA
C22.2 NO 601.1, ASTM F 2196-2002
IEC 60601-1-2