101515 REV C
1-19
Cleaning
INSTITUTIONAL (HOSPITAL) CLEANING INSTRUCTIONS
CAUTION
•
The use of quaternary disinfectant products may cause damage to the product.
•
Frequent or prolonged exposure to higher concentration disinfectant solutions may result in damage to the product.
Note
: Change the water monthly or more often depending on use.
Clean the outer surfaces of the T/Pump with one of the following:
•
A damp cloth and soapy water.
•
Phenolic Disinfectant or 10% bleach solution.
Note:
If water is contaminated or not changed for a month, please follow the internal cleaning procedure below.
1. Prepare a germicidal solution according to the manufacturer’s instructions. Stryker, product catalog MTA33.
2. Drain the pump.
3. Connect hose set together, where applicable
, or attach a pad to the connector hose.
4. Fill the reservoir to the
Heating
water line on the back of the reservoir.
5. (TP650) Select
Low Heat
on the keypad.
(TP700) Select the
95°F (35°C)
temperature setpoint on the keypad.
6. Start the T/Pump, and circulate the solution for one hour.
7. Drain the solution and refill the pump with clean water.
Note
: In a home environment, perform only step 2 and the refill instructions in step 7.
PADS / ACCESSORIES
Only use Mul•T•Pads
®
. The unique button design allows optimal water flow and provides trouble-free operation when
the pad is folded. This reduces the number of different sizes of pads your facility must keep in inventory.
The Mul•T•Pads with Clik-Tite
®
connectors can be interconnected to provide therapy to more than one body site at
a time (Figure 1B,
). Refer to Catalog Descriptions on
for a list of various pads and ordering
information.