Guidance and manufacturer's declaration - electromagnetic immunity
The System 6 handpiece is intended for use in the electromagnetic environment specified below. The customer or the user of the
System 6 handpiece should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment - guidance
Conducted Rf
IeC 61000--6
Vrms
150 kHz to 80 MHz
n/a
n/a
Portable and mobile Rf communications equipment
should be used no closer to any part of the System 6
handpiece, including cables, than the recommended sep-
aration distance calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance
d=1.67√
P
d=1.67√
P
80 MHz to 800 MHz
d=2.√
P
800 MHz to 2.5 GHz
Where
P
is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and
d
is the recommended separation
distance in meters (m)
Interference may occur in the vicinity of equipment
marked with the following symbol:
Radiated Rf
IeC 61000--
V/m
80 MHz to 2.5 GHz
V/m
80 MHz to 2.5 GHz
NOTe 1: At 80 MHz and 800MHz the higher frequency range applies.
NOTe 2: These guidelines may not apply in all situations. electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.