6205-001-700 Rev-A
2008/01
www.stryker.com
5
Figure 2 To Install Battery Pack
Instructions
To Install Battery Pack
NOTES:
• This handpiece is designed to accept a Stryker Large
or Small Battery Pack only (Ref 6215 or Ref 6212).
These battery packs can be charged in the Stryker
System 6 Battery Charger Ref 6110-120 configured
with the appropriate battery charger modules.
• See the instructions supplied with the battery charger
and/or battery pack for charging details and specifi-
cations.
Slide a fully charged battery pack firmly into the
handpiece until the battery latch snaps, indicating the
battery pack is secure (see figure 2).
1.
Test the operation of the handpiece by sliding the
f/R/Safe control to fORWARD (f) or fORWARD/Re-
VeRSe (f/R) and squeezing the trigger.
Slide the f/R/Safe control to the safe position until
you are ready to use the handpiece.
2.
.
Slide the f/R/Safe control to the safe position.
To install an attachment, align the driveshaft of the
attachment with the handpiece spindle. If the attach-
ment has notches, also align the notches with the
tabs in the sleeve of the handpiece (see figure 1).
1.
2.
Once properly aligned, insert the attachment into the
handpiece until the attachment “snaps” into place.
Tug the attachment to ensure it is secure.
Install the desired cutting accessory. See the
System
6 Rotary Handpiece Attachments
booklet.
.
.
5.
NOTE:
Several attachments are available and each
has a specialized accessory retainer.
WARNING:
To prevent inadvertent run-
ning of the handpiece, ALWAYS place the
f/R/Safe control in the safe position before
installing or removing any attachment.
To Install Attachment
Figure 1 To Install Attachment
Driveshaft
Tabs
Notches
(not always