6205-001-700 Rev-A
2008/01
www.stryker.com
Cleaning Recommendations
WARNINGS:
• Clean and sterilize handpieces, attachments and bat-
teries before first and every use.
• Prior to cleaning and sterilization, remove all attach-
ments and accessories from the handpieces.
• DO NOT use solvents, lubricants, or other chemicals
unless otherwise specified.
CAUTIONS:
• DO NOT immerse a handpiece, attachment or battery
pack in liquid. Moisture may enter the equipment,
cause corrosion, and damage the electrical and/or
mechanical components.
• DO NOT allow liquid to run directly into any electrical
connection. Moisture may cause corrosion to electri-
cal and/or mechanical components.
To Drain Water From Handpiece
To Clean Battery Packs and Accessories
See the care instructions supplied with the battery
packs, battery pack modules and battery charger.
To Clean Handpiece and Attachments
Remove the battery pack and attachment from the
handpiece.
Using a brush with stiff, non-metallic bristles and
hospital enzymatic cleaner, scrub the debris from the
handpiece and attachments. Pay special attention
to crevices and other hard to reach areas such as
seams, joints, and details around the retainer, trigger,
and connector areas. Use a bottle brush to clean the
cannula.
Rinse all external surfaces of the handpiece under
tap water. Hold the handpiece upright to prevent wa-
ter from running into the battery contact area. flush
the cannula with running water.
If water leaks into the handpiece, tip the handpiece
back as shown below to allow drainage from a small
opening in the battery pack contact area.
1.
2.
.
.
Visually inspect the handpiece for any remaining
debris; if any debris is present, repeat the cleaning
and rinsing procedure using fresh hospital enzymatic
cleaner.
Dry the handpiece and attachments with a lint-free
towel.
After cleaning, sterilize as directed. See
Sterilization
Recommendations.
5.
6.
7.