Troubleshooting Guide*

PROBLEM

CAUSE

ACTION

Handpiece does not run or 
turns at a reduced speed.

Battery pack is discharged.

Recharge the battery pack in Stryker battery 
charger.

Battery pack is expended.

Replace the battery pack.

f/R/Safe control is in the safe position.

Slide f/R/Safe control to the forward or 
forward/reverse position.

Drivetrain is malfunctioning.

Return the handpiece for repair.

Motor runs but cutting acces-
sory does not move.

Drivetrain is malfunctioning.

Return the handpiece for repair.

Attachment is not fully seated.

Remove and insert the attachment. ensure 
the attachment is fully seated.

Shifter knob is positioned between the DRILL 
and ReAM selections. 

Position shifter knob to either the DRILL or 
ReAM position.

Battery pack becomes unusu-
ally hot during use.

Circuitry is malfunctioning.

Check the battery pack on the charger.  
Replace the battery pack if required. See the 
instructions supplied with the battery charger.

Attachment will not fit into the 
handpiece.

Debris is on the attachment or inside the front 
end of the handpiece.

Clean the attachment and/or handpiece with 
a small brush with stiff, non-metallic bristles.

Attachment is damaged.

Return the attachment for repair.

Handpiece is damaged.

Return the handpiece for repair.

Attachment and/or cutting ac-
cessory wobbles in handpiece. 

Cutting accessory or wire/pin is bent, extends 
too far from the distal end of attachment, is 
the wrong size, or is not properly centered in 
the attachment.

Reinsert the cutting accessory, wire or pin. 
If wobble persists, return the handpiece and 
attachment for repair.

A reamer is installed and the handpiece is in 
the DRILL mode.

ALWAYS operate reamers in the ReAM mode.

*DO NOT service this equipment. If you require service, contact your Stryker sales representative or call Stryker customer service 
at 1-800-25-210. Outside the US, contact your nearest Stryker subsidiary.

Содержание System 6 6205

Страница 1: ...2008 01 6205 001 700 Rev A www stryker com Instructions For Use ref 6205 System 6 Dual Trigger Rotary Handpiece ...

Страница 2: ...ssories unless other wise specified Other accessories may re sult in increased electromagnetic emissions or decreased electromagnetic immunity of the system DO NOT modify any component or accessory Failure to comply may result in patient and or health care staff injury Important Information The words WARNING CAUTION and NOTE have special meaning and should be reviewed WARNING Disregarding WARNING ...

Страница 3: ...ipment like the handpiece Install and place the handpiece into service according to the EMC information in this manual Portable and mobile RF communications equipment can affect the function of the handpiece ALWAYS place the handpiece in the safe mode while the handpiece is idle before installing or removing any accessory or when passing the handpiece to an other person Failure to comply may resul...

Страница 4: ...REAM slow or DRILL fast rotate the knob to the appropriate position as required Accessory Interface Connector provides power and communication for future accessories Symbol Definitions FORWARD Fully depress the F R Safe control to lock the reverse trigger The forward trigger is functional REVERSE Press trigger to operate in the counterclock wise direction REAM For slow speed rotate the shifter kno...

Страница 5: ...ezing the trigger Slide the F R Safe control to the safe position until you are ready to use the handpiece 2 Slide the F R Safe control to the safe position To install an attachment align the driveshaft of the attachment with the handpiece spindle If the attach ment has notches also align the notches with the tabs in the sleeve of the handpiece see figure 1 1 2 Once properly aligned insert the att...

Страница 6: ...e F R Safe control to the safe position when you are finished operating the handpiece 1 2 CAUTIONS DO NOT stall the handpiece Failure to comply may damage the electric motor and or battery pack If the handpiece jams release the trigger immediately Remove any obstructions before continuing the pro cedure If any power loss is experienced while using a hand piece ALWAYS replace the battery pack with ...

Страница 7: ...he battery pack on the charger Replace the battery pack if required See the instructions supplied with the battery charger Attachment will not fit into the handpiece Debris is on the attachment or inside the front end of the handpiece Clean the attachment and or handpiece with a small brush with stiff non metallic bristles Attachment is damaged Return the attachment for repair Handpiece is damaged...

Страница 8: ...ng properly Ensure that there are no loose or missing components Check all moving parts for free movement Be alert for unusual sounds or vibrations and note the operating speed PROBLEM CAUSE ACTION Handpiece has become noisy and vibrates Drivetrain is malfunctioning Return the handpiece for repair Sporadic electrical interference is experienced Electrical noise is present Turn off all electrical e...

Страница 9: ... Handpiece and Attachments Remove the battery pack and attachment from the handpiece Using a brush with stiff non metallic bristles and hospital enzymatic cleaner scrub the debris from the handpiece and attachments Pay special attention to crevices and other hard to reach areas such as seams joints and details around the retainer trigger and connector areas Use a bottle brush to clean the cannula ...

Страница 10: ...posure time 8 minute minimum dry time To Sterilize Battery Packs See the care instructions supplied with the battery packs To Sterilize Handpieces and Attachments Sterilization Recommendations ETO 100 ETO 120 135 F 49 57 C Wrapped in an instrument tray or fully perforated sterilization box 2 hour 30 minute exposure time 8 hour minimum aeration time 250 F Gravity Gravity displacement sterilizer 250...

Страница 11: ...Intermittent Operation 1 minute on 4 minutes off 3 times with 3 hour rest Approval CSA International CAN CSA C22 2 No 601 1 M90 UL 60601 1 IEC 60601 1 Equipment Type Type BF Applied Part Power Supply Internally Powered 9 6 V Enclosure Protection IPX0 Ordinary Equipment Environmental Conditions Operation Storage and Transportation Temperature Relative Humidity Atmospheric Pressure Specifications ar...

Страница 12: ...ssions CISPR 11 Group 1 The System 6 handpiece uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B The System 6 handpiece is suitable for use in all establishments includ ing domestic establishments and those directly connected to the public low voltage power ...

Страница 13: ...han the recommended sep aration distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance d 1 67 P d 1 67 P 80 MHz to 800 MHz d 2 33 P 800 MHz to 2 5 GHz Where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in meters m Interference ma...

Страница 14: ...erial the relative humidity should be at least 30 Electrical fast transient burst IEC 61000 4 4 2 kV for power supply lines 1 kV for input output lines n a n a Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode n a n a Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle n a 40 UT 60 dip in UT for 5 cycles n a 7...

Страница 15: ... output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 n a 0 12 0 23 0 1 n a 0 37 0 74 1 n a 1 17 2 33 10 n a 3 70 7 37 100 n a 11 70 23 30 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable t...

Страница 16: ...ts 4100 E Milham Kalamazoo Michigan USA 49001 1 269 323 7700 1 800 253 3210 Stryker France ZAC Satolas Green Pusignan Av de Satolas Green 69881 MEYZIEU Cedex France ES DE FR IT NL JA ZH KO SV DA FI PT NO PL EL 6205 001 710 6205 001 720 6205 001 730 6205 001 750 ...

Отзывы: