C
Clle
ea
an
niin
ng
g
W
WA
AR
RN
NIIN
NG
G -- Always use any appropriate personal protective equipment while power washing to avoid inhaling contagion.
Power washing equipment may aerate contamination.
C
CA
AU
UT
TIIO
ON
N
•
Do not steam clean or ultrasonically clean the product.
•
Do not exceed 180 °F (82 °C) as the maximum water temperature.
•
Do not exceed 240 °F (115 °C) as the maximum air dry temperature (cart washers).
•
Do not exceed 1500 psi (130.5 bar) as the maximum water pressure. If you are using a hand held wand to wash the
product, keep the pressure nozzle at a minimum of 24 in. (61 cm) from the product.
The product is power washable. The product may show some signs of oxidation or discoloration from continuous washing.
No degradation of the product’s performance will occur from power washing as long as you follow the proper procedures.
•
Follow the cleaning solution manufacturer’s dilution recommendations exactly.
•
Power wash with recommended cleaners. Hose down the product and towel dry the guide.
•
Power wash the rails and interface plates with a hand held wand unit or wipe the product with a clean cloth and
recommended cleaners.
S
Su
ug
gg
ge
esstte
ed
d cclle
ea
an
ne
errss
In general, when used in concentrations recommended by the manufacturer, either phenolic type or quaternary (excluding
V
Viirre
exx® TB) type disinfectants can be used. Iodophor type disinfectants are not recommended for use because staining
may occur.
W
WA
AR
RN
NIIN
NG
G -- Always wipe the product with clean water and dry after cleaning. Some cleaning products are corrosive in
nature and may cause damage to the product. Failure to properly rinse and dry the product leaves a corrosive residue on
the surface of the product and may cause premature corrosion of critical components.
Suggested cleaners include:
•
Quaternary cleaners (active ingredient - ammonium chloride)
•
Phenolic cleaners (active ingredient - o-phenylphenol)
•
Chlorinated bleach solution (5.25% - less than 1 part bleach to 100 parts water)
Avoid over saturation. Do not allow the product to stay wet longer than the chemical manufacturer's guidelines for proper
disinfecting.
N
No
otte
e -- Failure to follow the above directions when using these types of cleaners may void this product’s warranty.
6373-009-005 Rev E.0
15
EN
Содержание M-1 6373
Страница 2: ......
Страница 4: ......
Страница 20: ......
Страница 22: ......
Страница 24: ......
Страница 42: ......
Страница 44: ......
Страница 62: ......
Страница 64: ......
Страница 82: ......
Страница 84: ......
Страница 102: ......
Страница 104: ......
Страница 122: ......
Страница 124: ......
Страница 140: ......
Страница 142: ......
Страница 144: ......
Страница 162: ......
Страница 164: ......
Страница 182: ......
Страница 184: ......
Страница 202: ......
Страница 204: ......
Страница 222: ......
Страница 224: ......
Страница 242: ......
Страница 244: ......
Страница 262: ......
Страница 264: ......
Страница 282: ......
Страница 284: ......
Страница 302: ......
Страница 304: ......
Страница 322: ......
Страница 324: ......
Страница 342: ......
Страница 344: ......
Страница 362: ......
Страница 364: ......
Страница 366: ......
Страница 384: ......
Страница 386: ......
Страница 402: ......
Страница 404: ......
Страница 406: ......
Страница 422: ......
Страница 424: ......
Страница 426: ......
Страница 443: ......
Страница 444: ...6373 009 005 Rev E 0 WCR AA 6 2019 08 ...
Страница 446: ......