P
Prro
occe
ed
du
urre
e::
1. Remove the foot support pivot assembly (see
Foot support pivot assembly removal and replacement
2. Insert the extension tube through the slide opening of the pivot assembly.
3. Using a 3/16” hex wrench, install the set screws into the bosses on the pivot.
4. Using a 3/16” hex wrench, alternately turn each set screw clockwise until the extension tube doesn’t slide.
5. Turn the assembly on end with the pivot assembly on top.
6. Using a 3/16” hex wrench, alternately turn each set screw counterclockwise until the pivot assembly slides down the
extension tube without any force applied.
7. Verify proper operation before you return the product to service.
F
Fo
oo
ott ssu
up
pp
po
orrtt sslliid
de
e -- ssiid
de
e//ssiid
de
e a
ad
djju
ussttm
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
3/16” hex wrench
P
Prro
occe
ed
du
urre
e::
1. Install four set screws in each side of the pivot.
2. Using a 3/16” hex wrench, turn each set screw clockwise until the extension tube is centered on the pivot assembly and
does not slide.
3. Turn the assembly on end with the pivot assembly on top.
4. Using a 3/16” hex wrench, alternately turn each set screw counterclockwise until the pivot assembly slides down the
extension tube without any force applied.
5. Verify proper operation before you return the product to service.
EN
24
1061-009-002 Rev A.0
Содержание Gynnie
Страница 2: ......
Страница 4: ......
Страница 36: ...EN 32 1061 009 002 Rev A 0 ...
Страница 45: ...L Liit tt te er r a as ss se em mb blly y 1061 030 010 Rev G Reference only 1061 009 002 Rev A 0 41 EN ...
Страница 46: ...EN 42 1061 009 002 Rev A 0 ...
Страница 77: ...Siderail pad set 1001 052 000 1061 009 002 Rev A 0 73 EN ...
Страница 83: ......
Страница 84: ...Stryker Medical 3800 E Centre Avenue Portage MI 49002 USA 1061 009 002 Rev A 0 WCR AA 10 2019 09 ...