Remove product from service before you perform the preventive maintenance inspection. Check all items listed during
annual preventive maintenance for all Stryker Medical products. You may need to perform preventive maintenance checks
more often based on your level of product usage. Service only by qualified personnel.
N
No
otte
e -- Clean and disinfect the exterior of the mattress before inspection, if applicable.
Inspect the following items:
All welds
All fasteners are secure
Brake mechanism works
Steer function works
Siderails raise, lower, and latch
Casters lock when you apply the brakes
Casters secure and swivel
Casters are free of wax or debris
Fowler backrest raises, lowers, and latches
Gatch raises, lowers, and latches
Skins are not cracked
Trendelenburg/Reverse Trendelenburg raises and lowers from all locations
IV pole is intact and adjusts and latches in all positions (optional)
Oxygen bottle holder is intact and works (optional)
No rips or cracks in the mattress cover
Body restraints work (optional)
Ground chain intact
No leaks at the hydraulic connections
Hydraulic jacks hold
Hydraulic oil level sufficient
Hydraulic drop rate is correct
Lubricate where required
Accessories and mounting hardware are in good condition
Product serial number:
Completed by:
Date:
EN
8
1061-009-002 Rev A.0
Содержание Gynnie
Страница 2: ......
Страница 4: ......
Страница 36: ...EN 32 1061 009 002 Rev A 0 ...
Страница 45: ...L Liit tt te er r a as ss se em mb blly y 1061 030 010 Rev G Reference only 1061 009 002 Rev A 0 41 EN ...
Страница 46: ...EN 42 1061 009 002 Rev A 0 ...
Страница 77: ...Siderail pad set 1001 052 000 1061 009 002 Rev A 0 73 EN ...
Страница 83: ......
Страница 84: ...Stryker Medical 3800 E Centre Avenue Portage MI 49002 USA 1061 009 002 Rev A 0 WCR AA 10 2019 09 ...