B
F
Fiig
gu
urre
e 5
5 –– B
Brra
akke
e a
ad
djju
ussttm
me
en
ntt
B
Ba
asse
e ssiid
de
e cco
on
nttrro
oll p
pe
ed
da
all lliin
nkka
ag
ge
e a
ad
djju
ussttm
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
3/8” hex wrench
•
7/16” wrench
•
5/32” hex wrench
•
Bungee cords
P
Prro
occe
ed
du
urre
e::
1. Place 150 to 200 lb on the litter.
2. Raise the litter to the highest position.
3. Raise the base hood and support the base hood with bungee cords.
4. Using a 3/8” hex wrench and a 5/32” hex wrench, loosen the pedal linkage adjustment nut until the jack stops drifting.
5. Raise the litter to the highest position.
6. Tighten the pedal linkage adjustment nut until the jack begins to slowly descend.
7. Using a 7/16” wrench, back the pedal linkage adjustment nut out four full turns.
8. Raise the litter to the highest position.
9. Loosen the pedal linkage adjustment nut and make sure that the jack does not drift. If the jack drifts, repeat this
procedure.
10. Verify proper operation before you return the product to service.
B
Brra
akke
e//sstte
ee
err ssiid
de
e cco
on
nttrro
oll g
ge
ea
arr rre
ep
plla
acce
em
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
1/8” hex wrench
1061-009-002 Rev A.0
13
EN
Содержание Gynnie
Страница 2: ......
Страница 4: ......
Страница 36: ...EN 32 1061 009 002 Rev A 0 ...
Страница 45: ...L Liit tt te er r a as ss se em mb blly y 1061 030 010 Rev G Reference only 1061 009 002 Rev A 0 41 EN ...
Страница 46: ...EN 42 1061 009 002 Rev A 0 ...
Страница 77: ...Siderail pad set 1001 052 000 1061 009 002 Rev A 0 73 EN ...
Страница 83: ......
Страница 84: ...Stryker Medical 3800 E Centre Avenue Portage MI 49002 USA 1061 009 002 Rev A 0 WCR AA 10 2019 09 ...