The Stryker cot fastener systems are compatible only with cots that conform to the installation specifications.
W
WA
AR
RN
NIIN
NG
G
•
Always install the cot fastener by qualified personnel only. Improper installation could result in injury to the patient or
operator.
•
Always make sure that all cots meet the installation specifications for the Stryker cot fastener system.
•
Always adjust the rail clamp assembly to match the cot retaining post position for the cot manufacturer and model
number.
These instructions are intended for cots with antler style cot fastener systems. For crash-rated cot fasteners, see the
appropriate Operations Manual for installation instructions.
IIn
nsstta
alllliin
ng
g tth
he
e iin
n--ffa
asstte
en
ne
err ssh
hu
utt--o
offff
These instructions are intended for cots with antler style cot fastener systems. For crash-rated cot fasteners, see the
appropriate Operations Manual for installation instructions.
W
WA
AR
RN
NIIN
NG
G
•
Always position the in-fastener shut-off before you place the cot into service.
•
Do not attempt to operate the cot when loaded into a cot fastener.
•
Always use the in-fastener shut-off to disable the electronic functionality only.
•
Always install the in-fastener shut-off system (if not using a crash-rated cot fastener) in any emergency vehicle that will
be used with this cot.
The cot and antler style cot fastener system have an integrated in-fastener shut-off function that disables the cot motor
when you secure the cot into the cot fastener. Tighten the bolts to secure the fastener before you install the shut-off bracket.
Install the shut-off bracket onto the rail clamp assembly before you place the cot into service.
1. Raise the base and push the cot into the vehicle patient compartment. Follow appropriate loading instructions.
2. Secure the extended head section of the cot into the cot fastener antler.
3. Secure the cot post into the fastener rail clamp.
4. Adjust the shut-off bracket along the rail clamp until the diamond (A) on the outer rail label is centered between the pop
rivet heads (B) on the in-ambulance shut-off (Figure 3).
5. Using a T27 Torx driver, securely fasten the bolts to attach the shut-off bracket to the rail clamp assembly.
6. Press the retract (-) button to check that the motor does not turn on while the cot is in the fastener. The cot battery LED
indicator will still illuminate. If the motor turns on, readjust the shut-off bracket.