Always place the electrical charger power cord where it will not be stepped on, tripped over, or otherwise subjected to
damage or stress.
•
Do not touch the battery receptacle terminals with metal objects.
•
Always grasp and pull the plug, not the cord, when you disconnect the charger to avoid the risk of damage to the
electrical plug and cord.
•
Always remove the battery before you wash the product.
•
Do not clean, service, or perform maintenance while the product is in use.
•
Do not steam clean or ultrasonically clean the product.
•
Do not exceed 180 °F (82 °C) as the maximum water temperature.
•
Do not exceed 1500 psi (130.5 bar) as the maximum water pressure. If you are using a hand held wand to wash the
product, keep the pressure nozzle at a minimum of 24 in. (61 cm) from the product.
•
Always use authorized parts to avoid the risk of product damage.
•
Do not lubricate the bearings in the X-frame as it will degrade the performance of the cot and may void its warranty.
•
Changes or modifications to the A
Allvva
arriiu
um
m Battery Management System, not expressly approved by Stryker, could void
the user's authority to operate the equipment.
EN
6
6507-009-001 Rev AB.3