![Stryker 6506 Скачать руководство пользователя страница 19](http://html.mh-extra.com/html/stryker/6506/6506_operation-manual_1385840019.webp)
www.stryker.com
6506-209-001 REV F
13
Introduction
Specifications - Power-PRO (Continued)
Charger
100-240 VAC 1.20 A,
50/60 Hz or 12 VAC 4.16 A -
SMRT
power system
Cot duty cycle
16.7% (less than 60 seconds on, more than 300 seconds
off)
Standards (cots and chargers)
ANSI/AAMI ES60601-1: 2012, CAN/CSA-C22.2 No. 60601-
1:14, KKK-A-1822
For standards that require specific options, see
Standards with required options on page 14
.
1
Cot loads over 300 lb (136 kg) may require additional assistance to meet the set cot load height.
2
Height measured from bottom of mattress at seat section to ground level.
3
Cot is weighed with one battery and without mattress and restraints.
4
Set the cot height to any ambulance deck height that ranges from 26 in. to 36 in. (66 cm to 91 cm).
Stryker reserves the right to change specifications without notice.
Power-PRO
XT is designed to conform to the Federal Specification for the Star-of-Life Ambulance (KKK-A-1822 ).
Power-PRO
XT is designed to be compatible with some competitive cot fastener systems.
The yellow and black color scheme is a proprietary trademark of Stryker Corporation.
Hereby, Stryker declares that the radio equipment type short range device is in compliance with Directive 2014/53/EU.
The full text of the EU declaration of conformity is available at the following internet address: http://techweb.med.
strykercorp.com/EMS/EU%20Declaration%20of%20Conformity/index.html.
Labels may be unreadable from a viewing distance greater than 12 inches.
WARNING
Power-PRO
with the
Power-LOAD
compatibility option operates primarily at these frequencies: 70 - 85 kHz for inductive
charging and 13.56 MHz±7 kHz, Amplitude Modulated (OOK), ERP: -82.37 dBm. The inductive charging can operate
between these frequencies: 70 - 125 kHz. Other equipment may interfere with the
Power-LOAD
system, even if that
other equipment complies with CISPR emission requirements.
Environmental conditions
Operation
Storage and transportation
Temperature
130 °F
(54 °C)
-30 °F
(-34 °C)
130 °F
(54 °C)
-30 °F
(-34 °C)
Relative humidity
93%
0%
93%
0%
Atmospheric pressure
1060 hPa
700
1060 hPa
700