3. Using a 1/8” hex wrench, remove the eight screws (F and G) that secure the column and drawer assembly to the top
cradle bracket (Figure 4).
F
Fiig
gu
urre
e 4
4 –– C
Co
ollu
um
mn
n a
an
nd
d d
drra
aw
we
err a
asssse
em
mb
bllyy
N
No
otte
e -- Lower the unit to the lowest height to help minimize the column dropping.
4. Using a T15 Torx driver, remove the four screws that secure both columns to each valance bracket.
N
No
otte
e -- The entire column and drawer assembly should now rest on top of the base cover.
5. Using a 1/8” hex wrench, remove the two screws (AL) (Figure 5) that secure each release handle to the top cradle
bracket. Save the handle and screws.
EN
8
4402-009-002 Rev AB.1