F
Fiig
gu
urre
e 1
13
3 –– T
To
op
p ccrra
ad
dlle
e b
brra
acckke
ett
N
No
otte
e -- Add loctite blue threadlocker to a minimum of three threads on item L.
4. Using a rubber hammer, strike the top cradle bracket, near the gas cylinder to remove the cradle bracket from the gas
cylinder.
5. Remove the cradle assembly and supports to gain access to the gas cylinder. Set aside the cradle assembly.
6. Lift up and remove the cabinet/drawer assembly. Set aside the cabinet/drawer assembly.
7. Using a 5/32” hex wrench, remove the two screws that secure the column support bracket. Save the screws.
8. Using a 3/16: hex wrench, remove the screw (AJ) and spacer (AH) from the lower guide tube (Figure 14). Save the
screw and spacer.
F
Fiig
gu
urre
e 1
14
4 –– C
Co
ollu
um
mn
n ssu
up
pp
po
orrtt b
brra
acckke
ett
9. Using the needle nose pliers, remove the gas cylinder clip. Save the clip.
10. Grab the shaft of the gas cylinder and pull up to remove the gas cylinder from the base support bracket. Save the
washers (D), bumpers (C) and clip (F) for reinstall (Figure 15). Discard the gas cylinder.
4402-009-002 Rev AB.1
17
EN
Содержание 4402-000-001
Страница 2: ......
Страница 38: ...EN 36 4402 009 002 Rev AB 1 ...
Страница 41: ......
Страница 42: ...Stryker Medical 3800 E Centre Avenue Portage MI 49002 USA 4402 009 002 Rev AB 1 WCR AA 6 2021 11 ...