![Stryker 4402-000-001 Скачать руководство пользователя страница 14](http://html.mh-extra.com/html/stryker/4402-000-001/4402-000-001_maintenance-manual_1385825014.webp)
5. Using a T15 Torx driver, remove the six screws (AP) and washers (AN) (Figure 8) that secure two valance brackets.
Save the screws and washers.
F
Fiig
gu
urre
e 8
8 –– V
Va
alla
an
ncce
e b
brra
acckke
ettss
6. Using a 1/2” speed ratchet remove the four screws (L) and washers (K) (Figure 9) that secure the push bar to the cradle
bracket. Save the screws and washers.
EN
12
4402-009-002 Rev AB.1
Содержание 4402-000-001
Страница 2: ......
Страница 38: ...EN 36 4402 009 002 Rev AB 1 ...
Страница 41: ......