3. Using a 3/8” drive ratchet and 5/16” hex driver, remove the bolt (N) and caster (K, L, M) (Figure 10) assembly from the
base weldment. Discard the caster assembly.
Torque item N
to 35 to 45 ft-lb
Torque item N
to 35 to 45 ft-lb
Torque item N
to 35 to 45 ft-lb
Torque item N
to 35 to 45 ft-lb
F
Fiig
gu
urre
e 1
10
0 –– B
Ba
asse
e w
we
elld
dm
me
en
ntt
4. Reverse to install.
5. Verify proper operation before you return the product to service.
D
Drra
aw
we
err sslliid
de
e rre
ep
plla
acce
em
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
T15 Torx driver
•
T10 Torx driver
•
5/16” Wrench
P
Prro
occe
ed
du
urre
e::
1. Apply the brakes.
2. Pull open the drawer.
3. Using a T10 Torx driver and 5/16” wrench, remove the two screws that secures the drawer slide to the drawer. Save the
screws.
4. Using a T15 Torx driver, remove the screw (E) and washer (F) (Figure 11) that secure the drawer slide to the column.
Save the screw and washer.