![Stryker 4402-000-001 Скачать руководство пользователя страница 12](http://html.mh-extra.com/html/stryker/4402-000-001/4402-000-001_maintenance-manual_1385825012.webp)
F
Fiig
gu
urre
e 6
6 –– C
Cyylliin
nd
de
err b
brra
acckke
ett
N
No
otte
e -- Add loctite blue threadlocker to a minimum of three threads on item L.
7. Using a rubber hammer, strike the top cradle bracket, near the cylinder to disengage the bracket from the cylinder top.
Repeat on opposite side.
8. Using a workbench or table, flip the top assembly upside down.
9. Using a T15 Torx driver remove the six screws (AP) and washers (AN) that secure the two valance brackets. Save the
screws and washers.
10. Working from one side of the assembly, firmly pull up on the cradle bottom cover.
11. Remove and discard the cradle bottom cover.
12. Reverse to install.
13. Verify proper operation before you return the product to service.
P
Pu
ussh
h b
ba
arr rre
ep
plla
acce
em
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
1/8” Hex wrench
•
T15 Torx driver
•
1/2” Speed ratchet
•
Loctite blue threadlocker
P
Prro
occe
ed
du
urre
e::
1. Apply the brakes.
2. Remove the cradle top cover. See
Cradle top cover removal/replacement
EN
10
4402-009-002 Rev AB.1
Содержание 4402-000-001