Service Information
23
Replacing Scale CPU:
Required Tools:
Phillips Head Screwdriver
Small Screwdriver
Replacement Procedure:
1.
Unplug the power cord from the wall receptacle.
2.
Properly ground yourself (see page 14 for static discharge precautions).
3.
Remove the foot end frame end from the litter (see page 16 for instructions).
4.
Remove the connectors from the scale CPU circuit board (see illustration). Note the orientation and
placement of the connectors so they will be attached properly to the new scale CPU board.
Scale Power Cable
Weigh Cable
Shield to Ground
Scale Cable Ass’y
Weigh System
Cable
JP2
JP1
JP3
JP5
JP4
5.
Using a small screwdriver, remove the four screws holding the scale CPU board and carefully remove
the CPU board from the spacers on the litter computer plate.
6.
Place the new scale CPU board over the spacers and reinstall the four screws.
7.
Attach the connectors to the scale CPU board. Be sure to orient the connectors properly.
Calibration Procedure:
1.
Place the new CPU board in diagnostics mode.
2.
Push the scale lockout toggle switch (located under the foot end control panel) to the off position.
3.
Press and hold the LBS./KG. button and push the toggle switch back to the on position.
4.
Release the LBS./KG. button. The scale display should read “DOWNLOAD BARCODE”.
5.
Repeatedly press and release the LBS./KG. button until the display reads “INIT TO DEFAULTS”.
6.
Press and hold the WEIGHT SYSTEM ON button until the display reads “DEFAULTS ENTERED”.
7.
Repeatedly press and release the LBS./KG. button until the display reads “CALIBRATE SCALE”.
8.
Press and hold the WEIGHT SYSTEM ON button. Zero the scale, following the displayed instructions.
When the scale is zeroed, the scale monitor will display “REF X 100 = <2>0000. This reference number
represents 200 pounds.
9.
If the weight being used for calibration is not 200 pounds, the reference number must be changed to
match the weight. To change the reference number, use the LBS./KG. button to position the brackets
(< >) around the digit to be changed and the LOCK/UNLOCK button to scroll through the numbers.
Содержание Renaissance Series
Страница 33: ...Base Assembly 32 Assembly part number 2020 1 210 reference only ...
Страница 37: ...Base Assembly with 110V Outlet Option 36 Assembly part number 2020 11 210 reference only ...
Страница 39: ...Base Hydraulics Assembly 38 Assembly part number 2020 70 210 reference only ...
Страница 40: ...Base Hydraulics Assembly 39 ...
Страница 49: ...International Hood and Base Assembly 48 N FUSE JUNCTION BLOCK CORD ...
Страница 52: ...Basic Litter Assembly 51 B Assembly part number 2020 1 513 reference only ...
Страница 53: ...Basic Litter Assembly 52 LITTER CONTROL BOARD CONNECTIONS MOTOR WIRE ROUTING E ...
Страница 55: ...Litter Assembly Basic Bed 54 CABLE ROUTING DETAIL Assembly part number 2020 49 510 reference only ...
Страница 57: ...Litter Assembly Position Option 56 CABLE ROUTING DETAIL Assembly part number 2020 50 510 reference only ...
Страница 64: ...Notes 63 ...
Страница 65: ...Litter Assembly All Options without Head End Keypad 64 Assembly part number 2020 53 510 reference only ...
Страница 68: ...Litter Assembly Scale Option 67 CABLE ROUTING DETAIL LITTER CONTROL BOARD WEIGH BOARD CONNECTIONS ...
Страница 115: ...2020 250 410 Integrated Dynamic Mattress System Assembly 114 ...
Страница 123: ...DH 2 00 2020 1 152 REV K 6300 Sprinkle Road Kalamazoo MI 49001 ...