XENON GLADIATOR III TROUBLESHOOTING
Bulb fails to ignite.
1.
MODE switch S3 set to AUTO. Place in MAN. position when not employing automated or remote
lamphouse operation.
2.
AC power not on to lamphouse. Turn switching power supply 30 A. circuit breaker ON. If 115 V.AC
not read at 2 & 4, or POWER indicator not glowing, see power supply manual.
3.
Door interlock switch (S1) open. Close and secure lamphouse door.
4.
Faulty interlock switch(s). Check for 115 V.AC at 9 & 12; replace switch(s) if defective.
5.
Air vane switches not closing. Check for unobstructed operation; clean if required. Check continuity
between NO and COM; replace if defective.
6.
Faulty S2 ON-OFF switch. Check for voltage at 3 & 5; check for loose terminals or wiring. Replace
if defective.
7.
Blower fuse F1 blown. Replace as required (1.5 A. Std.).
8.
Low AC source voltage actuating brownout protection circuit in xenon power supply.
Bulb fails to ignite; ping audible, bulb flash visible.
1.
Inadequate DC output from xenon power supply. Set power supply output to correct range required for
bulb wattage (95 A. nominal for 3000 watt).
2.
If bulb flash is visible but faint, check for defective HV capacitor(s) in igniter. Replace if defective.
3.
Faulty or expired xenon bulb. Replace as required.
Bulb fails to ignite; ping audible, no bulb flash.
1.
Faulty xenon bulb. Check for cracked electrodes or darkened envelope. Replace if defective.
2.
Ignition pulse shorting to ground. Inspect DC leads for burned insulation; dress leads away from
grounded metal components.
No high voltage ping audible; MODE switch in MAN. and LAMP switch in ON.
1.
Loss of AC control voltage. Check xenon power supply for tripped circuit breaker or open thermal
switch. See power supply manual.
2.
Little or no DC No Load voltage. Measure DC No Load voltage at 10 & 15. See power supply
manual.
3.
Open fuse F1 (600 V.) on switching power supply. SEE POWER SUPPLY MANUAL. Allow (20)
minutes for capacitor discharge before replacing.
4.
Faulty igniter printed circuit board. If lamphouse ignites immediately when emergency ignition switch
is pressed, replace printed circuit board.
5.
Loose spark gap connections or terminals. Repair or replace as required.
7.
Faulty igniter. Check for 115 V.AC at 5 & 6; adequate DC No Load at 10 & 15. If present, and
igniter does not fire, replace igniter.
XG3/022
Содержание XENON GLADIATOR II
Страница 2: ......
Страница 17: ...XG3 015 ...
Страница 27: ...FIGURE 1 XG3 025 1 2 3 4 5 6 7 8 9 10 11 12 ...
Страница 29: ...FIGURE 2 XG3 027 1 2 3 4 5 6 7 8 9 10 11 12 ...
Страница 31: ...FIGURE 3 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 XG3 029 ...
Страница 34: ...FIGURE 4 XG3 032 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 ...
Страница 36: ...1 2 3 4 5 6 7 8 9 10 FIGURE 5 IGNITER ASSEMBLY XG3 034 ...
Страница 39: ...FIGURE 6 XG3 037 ...
Страница 41: ...FIGURE 7 XG3 039 ...
Страница 43: ...FIGURE 8 XG3 041 ...
Страница 46: ...58 59 FIGURE 9 Item 58 Bracket mounts to back surface of Item 31 XG3 044 ...
Страница 50: ...FIGURE 10 47 XG3 048 ...
Страница 53: ......
Страница 54: ......
Страница 55: ......