Errors and messages
Instructions for use • UNIDRIVE Select • PCC270_EN_V1.0_02-2022_IFU_CE-MDR
60
10 Errors and messages
10.1 Troubleshooting
Fault
Possible causes
Actions
Application instrument not
functional
Power supply failure
Check power supply
Product failed
Contact Service
Power switch lamp does not
light up
Power supply failure
Check power supply
Defective fuse
Changing a fuse
Power switch lamp defective
Contact Service
Power switch lamp lights up:
–
No rotation
–
No reaction when
footswitch is pressed
–
No reaction when
tapping on the touch
screen
–
Speed cannot be
regulated
Faulty motor control electron-
ics
Contact Service
Connection to motor inter-
rupted
Faulty handpiece
Connect another
handpiece
Handpiece blocked or
jammed
Clean handpiece
Tubing set connectors do not
fit in the pump
Tubing set installed incor-
rectly
Check the connection
side (upper and lower
connectors are different
sizes)
Connectors secured incor-
rectly to the tube
Replace tubing set
No irrigation fluid
Irrigation bag empty
Use new irrigation liquid
bag
Tubing clamp impedes flow
Loosen tubing clamp
Kinked tubing
Change run of tubing
Irrigation bottle not ventilated
Ventilate the irrigation
bottle
Pump deactivated
Activate pump
10.2 Software messages
Message
Possible causes
Actions
Motor not supported
Motor cannot be used with
available license
–
Motor defective
Connector problems
Contact Service
Product defective
Motorboard defective
Contact Service
Footswitch defective
Pedal defective
Contact Service
Pedal cannot be calibrated
Содержание UNIDRIVE Select
Страница 1: ...Instructions for use UNIDRIVE Select en...