Safety and warning
Instructions for use • UNIDRIVE Select • PCC270_EN_V1.0_02-2022_IFU_CE-MDR
9
systems. Local regulations take precedence over IEC and ISO standards. If you have any
questions, please contact your local KARL STORZ representative or technical customer
support.
Only combine the product with devices and components that are approved for combined
use by the manufacturer.
Only use devices and components that have standardized interfaces and do not breach
the intended use of the product.
Connect only devices that comply with IEC 60601 to signal inputs and signal outputs.
Observe the instruction manuals and interface specifications of the devices and
components used in combination.
Comply with national and local regulations.
Only make changes to the product if these changes are approved by KARL STORZ.
3.5 Risk of explosion and fire
The product can generate sparks, which cause combustible or flammable gases and liquids to
ignite or explode. This may cause injuries to patients, users, and third parties.
When using explosive narcotic gases: Operate the product outside of the hazard zone.
Do not use the product in the presence of flammable anesthetics.
The product must not be operated in oxygenated environments.
Only connect or disconnect the power plug to or from the power supply outside explosive
atmospheres.
3.6 Dangers from electrical current
An improper power supply may cause an electric shock and injure patients, users, or third
parties. All electrical installations of the operating room in which the device is connected and
used must meet the applicable IEC standards.
Use either the power cord supplied by KARL STORZ or a power cord which has the same
properties and which bears a national mark of conformity.
The product may only be operated with the line voltage stated on the rating plate.
Position the product appropriately so that the power cord can be unplugged at any time.
The product is only voltage-free when the mains plug has been disconnected.
The product’s ground line should be installed by a qualified electrician.
Ensure potential equalization according to the applicable national rules and regulations.
To ensure reliable protective earth grounding, connect the product to a properly installed
socket that is approved for use in the operation room.
Connect the product to a power supply with protective conductor.