13 Technical data
ASTOTHERM PLUS Instructions for Use
47
13 Technical data
ASTOTHERM PLUS
REF
AP220..
..AU
..EU
..UK
..NA
Electrical connection
230 - 240 VAC
50
– 60 Hz
100 - 115 VAC
50
– 60 Hz
Fuses primary (F1 F2)
T4AH 250 V (5 x 20 mm)
Fuses secondary (F3 F4)
T0630AL 250 V (TR5 type)
Power consumption
max. 450 W
Classification (IEC/EN 60601-1)
Class I, Type B
applied part is defibrillation proof
Classification (IEC/EN 60529)
IPX4
Classification (MDD 93/42/EEC)
Class IIb
Code UMDNS
10-447
Code GMDN
47616
Regulatory class as per FDA
II
Dimensions
Height
Width
Depth (incl. attachment device)
max.
145 mm
135 mm
295 mm
Weight
2.9 kg
Operating mode
continuous operation
Permissible environmental
conditions
Humidity
Temperature
Pressure
in operation
10% to 90%
not condensing
+16°C to +32°C
700 hPa to
1060 hPa
In storage
10% to 90%
not condensing
-20°C to +60°C
500 hPa to
1060 hPa
Selectable setpoint temperatures
Standard setting by manufacturer
41°C
39°C
37°C
Essential performance acc. to
IEC/EN 60601-1
Regulation of the temperature of the heat
exchanger to a selectable Set-temperature between
37 °C to 41°C +/- 0.5 °C
1.Excessive temperature cut off
2. Excessive temperature cut off
42.5°C (± 0.5°C)
43.5°C (± 0.5°C)
Low temperature alarm
T
Setpoint
– 4°C (± 0.5°C)
Heating up time (22°C to 40°C)
approx. 1 minute
Self-start
after power interruption up to
5 seconds