ASTOTHERM PLUS Instructions for Use
6 Product description
18
6.4 Control panel
Fig. 3: Control panel
#
Item
Description
1
“Standby” LED
Illuminates blue when the warmer is in
Standby Mode
.
2
“Standby” button”
Turns the warmer from
Standby Mode
to
On Mode
.
Turns the warmer from any
Mode
to the
Standby Mode
.
3
“On” LED
Illuminates green when the warmer is in
On Mode
.
4
“Start” button
Starts the heating process while the device is in
On Mode
or
Alarm Mode
.
Starts test 6 (see
chapter 12.1 Recurrent tests
), when the warmer is
operated with the middle setpoint temperature.
5
“Start” LED
Flashes green when the device is in
On Mode
(heating is not
yet started).
Illuminates green when the device is in
Heating Mode
(
“Start”
button has been pressed).
6
“Decrease” button
Selection of the next lower setpoint temperature. The frame
indicates the selected temperature.
Starts test 8 (see
chapter 12.1 Recurrent tests
), when the warmer is
operated with the lowest setpoint temperature.
7
“Increase” button
Selection of the next higher setpoint temperature. The frame
indicates the selected temperature.
Starts test 7 (see
chapter 12.1 Recurrent tests
), when the warmer is
operated with the highest setpoint temperature.