ASTOTHERM PLUS Instructions for Use
12 Maintenance
46
Test 5:
Temperature control and display
Value [°C] Min [°C] Max [°C]
P/F
Selected setpoint temperature
T
(max. 41°C)
Temperature measurement
TT
(with clinical thermometer)
T
– 0.5
T
+ 0.5
Actual temperature (displayed)
TD
T
– 0.3
T
+ 0.3
Manual tests
P/F
Test 6:
Manual excessive temperature cut off 1 (t1)
Test 7:
Manual excessive temperature cut off 2 (t2)
Test 8:
Manual low temperature alarm (t3)
Overall assessment
Please mark where applicable
No safety or functional deficiencies were detected
No direct risk, deficiencies detected may be corrected on short term
Equipment shall be taken out of operation until deficiencies are corrected
Equipment does not comply
– Modification / Exchange of components / Taking
out of service - is recommended
Comments
Date
Signature