Instruction Manual for Sicuro pesa Hospital Bed
Page
85
Disposal of gas springs/hydraulic units
Any gas springs and hydraulic units are primarily constructed from metal and
plastic and can be recycled.
Before disposing of these according to the manufacturer's
instructions, it is important to drain off the oil and dispose of it
properly.
Please note in this connection:
The release mechanism must not be activated if gas springs are
removed. These devices are under pressure.
Careless release could cause injury!
-
Gas springs must first be depressurised according to the
manufacturer’s instructions before disposal. This information
can be obtained upon request from the gas spring
manufacturer (see type plate).
In other countries outside Germany or the EU, the relevant national regulations
must be complied with.