Instruction Manual for Sicurao pesa Hospital Bed
Page
30
4.2.4.6
Lateral Tilting
(Optional Feature)
The mattress base of this hospital bed can be tilted sideways along its longitudinal
axis to an angle of approximately 25°. For safety reasons, this adjustment can only
be carried out from the control box.
Before tilting the bed along its longitudinal axis, you MUST protect
the patient from falling out of bed!
Use safety sides, shaped cushions (accessory) or similar
preventive measures for this purpose.
For safety reasons, this function must be kept locked at all times
(display
lights up orange) while not in use. This prevents
inadvertent adjustments from being made and the patient from
falling out of bed.
Enabling the function
.
By pressing the
or the
button and a release button
, the mattress
base of the bed will tilt along its longitudinal axis.
While the mattress base is being tilted, the motors will stop when the
horizontal position is reached even if the buttons are still depressed. In
order to continue tilting beyond the horizontal position, the buttons must
first be released and then pressed again. This makes it easier to adjust
the bed to a horizontal position.
The mattress base can be tilted while at any height. If necessary, the
height will be automatically adjusted.
Tilting to the left
Tilting to the right
Enabling/locking the tilting
function: