Instruction Manual for Sicurao pesa Hospital Bed
Page
72
7.2
E
RROR
C
ODES
D
ISPLAYED
For safety reasons, if a serious error is detected by the control unit CB16, the
functions concerned are electronically locked.
The various faults below are displayed by the locking LED flashing and this makes
it easier to rectify the fault:
Flashing locking LEDs on:
Control Box
Possible Faults
Solution
A
B
C
D
E
2 1 2 1 2 1 2 1 2
Position switch for mattress
base height motor M1
,
head end (socket 1)
Check cable + plug
Replace motor
RESET
Position switch for
mattress base height
motor M3
, foot end
(socket 3)
Position switch for
mattress base height
motor M4
, lateral tilting,
(socket 4)
Handset fault, internal
control fault
Check cable + plug
Replace handset
Replace control unit
RESET
1
A
2 1
B
2 1
C
2 1
D
2 1
E
2