Instruction Manual for Sicuro pesa Hospital Bed
Page
65
6.5.2.3
Initialising Drives (Alignment)
Version EM4
Version EM5
Where?
On the control box [
13
] at the foot end.
1.
Press and hold both
"RESET"
buttons
at the same time
2.
An
acoustic signal will sound.
3.
Release the buttons as soon as the signal stops after 5 seconds.
The control unit is now in
manual mode
:
The drives which require to be re-adjusted must now be moved to their end
position (=reference position).
o
Only necessary for: Drives M1 + M3 (mattress base height); M4 (lateral
tilting)
Manual mode is exited automatically if:
o
No button is pressed for 20 seconds, or
o
All drives have been re-initialised
While the drives to be adjusted are being moved, a pulsating beep is
emitted to indicate "drive not yet initialised" - control unit is in "manual
mode".
In manual mode, there is no collision monitoring facility activated -
please bear this in mind when moving the various drives to prevent
damage to the bed caused by the mattress base colliding with the
chassis or the crank overturning in the chassis.
1