Instruction Manual for Sicurao pesa Hospital Bed
Page
54
A distance of 30 cm between the spray valves and electrical
components has to be ensured. The only spray nozzles which are
permissible are flat jets.
During the washing and disinfection procedure (including rinsing),
the bed temperature may not exceed 70
C. Washing
temperatures that are too low must also be avoided as this will
result in poor drying.
Max. wash water temperature 65 °C!
Do not cool suddenly using cold water.
After completing the washing procedure, the bed must cool down
for an appropriate period of time (10 to 20 minutes) before it may
be put into service again.
The draining and drying processes can be optimised by:
Setting the mattress base of the bed to a reverse-Trendelenburg
position (approx. 10°)
Raising the safety sides on one side only