Installing the Monitor
Vividimage
®
4K Surgical Display
OPM3000, Rev. C
Operation Manual
8
Installing the Monitor
Here's how to install the Vividimage
®
4K Surgical Display, shown below.
1.
Unpack the monitor and other items from the carton.
If any of the items are missing (see “Monitor and Accessory Checklist” on page 3), please call Customer
Support at 800-548-4873.
2.
Attach the monitor to the mount with the included screws.
Note:
If the display will be mounted on a boom or Equipment Management System, the boom vendor
must wire the appropriate cables through the boom arm..
Three sets of screws come with your monitor: four pan-head Phillips 16mm screws, four pan-head Phillips
10mm screws, and four 16mm flat-head socket screws. You need only four screws to attach the monitor. For
mounts under 5mm (0.2in) thick, use the shorter pan-head Phillips screws. For mounts with thicker mounting
plates, use the longer pan-head Phillips screws. For the HarmonyAIR yoke, which has countersunk holes in its
mounting plate, use the (4) four M4 x 18mm flat-head socket screws. If these screws aren't available, use
screws that are 4.5mm to 9mm (0.18in to 0.35in) longer than the mount's thickness.
WARNING:
Be sure to use screws of the recommended size. If you use longer screws, you could
damage the internal workings of the monitor. If you use smaller screws, the monitor may not be securely
fastened.
Insert the screws in one of the three hole patterns shown in Figure 2. The hole patterns are compatible with
mounts that use the VESA (Video Electronics Standards Association) 100mm x 100m mounting pattern. The
mount is sold separately.