Table 2
(Page 24, Table 2 from IEC 60601-1-2:2014)
Vividimage
®
4K Surgical Display guidance and manufacturer's declaration - electromagnetic immunity
The Vividimage
®
4K Surgical Display is intended for use in the electromagnetic environment specified below.
The customer or the user of the Vividimage
®
4K Surgical Display should assure that it is used in such an environ-
ment.
Immunity Test
IEC 60601 test level
Compliance level
Electromagnectic environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
Complies
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/out-
put lines
Complies
Mains power quality should be that of a typ-
ical commercial or hospital environment.
Surge
IEC 61000-4-5
±1 kV differential
mode
±2 kV common
mode
Complies
Mains power quality should be that of a typ-
ical commercial or hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
<5%
U
T
(>95% dip in
U
T
)
for 0,5 cycle
40%
U
T
(60% dip in
U
T
)
for 5 cycles
70%
U
T
(30% dip in
U
T
)
for 25 cycles
<5%
U
T
(>95% dip in
U
T
)
for 5 sec
Complies
Mains power quality should be that of a typ-
ical commercial or hospital environment. If
the user of the Vividimage
®
4K Surgical Dis-
play requires continued operation during
power mains interruptions, it is recom-
mended that the Vividimage
®
4K Surgical
Display be powered from an uninterruptible
power supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
Complies
Power frequency magnetic fields should be
at levels characteristic of a typical location
in a typical commercial or hospital environ-
ment.
NOTE:
U
T
is the AC mains voltage prior to application of the test level.