Storage & Installation
•
Prior to installation, store in a dry location in temperatures between -4º Fahrenheit (F) and 140º F / -20º Celsius
(C) and 60º C and a barometric pressure of 700 hPa to 1060 hPa.
•
Always following installation instructions to avoid physical injury or damage to the device.
•
Always follow mounting instructions to avoid physical injury and/or damage to the monitor.
•
Take care when lifting this device. Use proper safety precautions to protect against injury.
•
The installation must be carried out in accordance with all applicable installation rules.
•
Do not place on unstable surfaces.
•
Do not modify or tamper with AC line.
•
Use of accessories, transducers and cables other than those specified or provided by the manufacturer of
this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity
of this equipment and result in improper operation.
Location
•
Use the monitor in a suitable environment. See “Operating Temperature” and “Storage Temperature” in
“Specifications” on page 18.
•
Do not place the display near active high-frequency medical electrical equipment such as magnetic
resonance image (MRI) devices where the electromagnetic interference may be high.
•
Do not install near heat sources such as radiators, heat registers, stoves, or other apparatus (including
amplifiers) that produce heat.
•
Even though the monitor is classified as surgical / medical grade, use caution around liquids as you would
with any electrical appliance.
•
Portable RF communications equipment (including peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm (12 inches) to any part of the STERIS Surgical Display
including cables specified. Otherwise, degradation of the performance of this equipment could result.
•
The equipment should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is
necessary, the equipment should be observed to verify normal operation in the configuration in which it will
be used.
•
Do not install this apparatus outdoors.
•
Do not insert objects into the monitor.
•
Do not place the monitor on unstable surfaces.
•
In all cases, refer to the specifications in this manual to ensure proper monitor performance. Use of the
monitor outside of operating specifications will void the monitor warranty and may cause permanent damage
to the monitor.
The CE marking certifies that this product has met European Union consumer safety, health or environ-
mental requirements
The European Authorized Representative (AR) for this product on behalf of the manufacturer is:
STERIS Ireland Limited
IDA Business and Technology Park
Tullamore
County Offaly
R35 X865
Ireland
Tel: +44(0)166 276 8636
24 Hr. Emergency No. +44(0)1895 622639