4-11
Processing Instructions
Operator Manual
920011-649
f.
While control handle is resting on control handle boot frame,
secure split insertion tube sleeve into control handle boot
opening (see Figure 4-6).
Verify split insertion tube sleeve handle is pushed down.
Insertion tube must be inserted over sleeve handle to avoid
.
g. Once split insertion tube sleeve is correctly positioned and
secured, finish inserting rest of insertion tube in tube slide
(see Figure 4-7).
13. Place endoscope control handle in control handle boot "A". Verify
light guide cable strain relief boot is placed between universal
cord port flaps (see Figure 4-7).
14. Place and coil light guide/ultrasound cables carefully in large
mesh basket (see Figure 4-11). Allow cables to coil naturally. Be
careful not to kink or overbend cable.
15. Rest light guide/ultrasound connectors in large mesh basket.
DO
NOT
place light guide/ultrasound connectors on top of light
guide/ultrasound cables (see Figure 4-11).
16. If processing a second endoscope, place it in control handle
boot "B", repeating
If only one endoscope is being processed
, insert control
handle boot plug into fluid delivery port of control handle boot
"B", as shown in Figure 4-12. Firmly insert control boot plug into
port. Close control handle boot "B".
17.
If processing an endoscope identified as requiring use of a Flow
Unit
, connect Flow Unit (refer to
Endoscope Quick Reference
Guide
).
Flow Unit
Connectors
Coil Light Guide
Cable
Channel
Cleaning Brush
Light Guide
Connector
Figure 4-11. Coil Light Guide Cable and Connect
Flow Units
Содержание P920011-649
Страница 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Страница 2: ......
Страница 13: ...xi 920011 649 Operator Manual Table of Contents ...
Страница 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Страница 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Страница 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Страница 167: ......
Страница 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...