4-4
920011-649
Operator Manual
Processing Instructions
then:
then:
and printer records, for example:
===========================
PROGRAM = 117-XXX-XXX
REV. = XX MM-DD-YY
===========================
CONTROL ON HH:MM:SS XM
MM-DD-YY
===========================
Reliance
Endoscope
Processing
System
S/N 36XXXXXXXXX
===========================
NOTE:
If display shows D-SHORT REQUIRED XX HOURS
LEFT or D-SHORT REQUIRED XX MINUTES LEFT, refer to
. If display shows D-LONG
REQUIRED NOW, refer to
2. Press
UNLOCK
toggle switch to unlock processor door. Open
door.
3. Manually rotate spray arms on chamber ceiling and under
endoscope processing support to ensure movement is not
obstructed. If spray arms do not rotate freely, refer to
4. Pull endoscope processing support out of chamber and onto
door.
5. Verify:
a. Chamber is empty and free of objects.
b. Removable debris screen at bottom of chamber is in place
and free of debris (refer to Figure 3-3).
6. Place accessories in Reliance EPS Processor (refer to
Endoscope Quick Reference Guide
for more details):
a. Place bite blocks on accessory pegs (see Figure 4-2).
b. Open accessory basket cover.
c. Place small items, such as air/water valves, in accessory
basket (see Figure 4-2, Detail A).
STERIS CORP
COPYRIGHT (C) 2011
READY
MM-DD-YY
HH:MM XM
WARNING – INEFFECTIVE
PROCESS HAZARD:
•
Follow
all
applicable
instructions provided by the
endoscope manufacturer for
preparing endoscope and
accessories for immersion.
Failure to do so may result
in damage to endoscope and/
or ineffective processing.
•
Before starting a cycle,
ensure rotary spray arms on
endoscope
processing
support and within chamber
are not obstructed and rotate
freely. If rotary spray arms
are obstructed, equipment
damage may result, and
processor will be unable to
effectively process load.
WARNING – INEFFECTIVE
PROCESS HAZARD: To avoid
equipment damage and help
ensure processing efficacy,
always follow directions when
loading items in endoscope
processing support.
Содержание P920011-649
Страница 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Страница 2: ......
Страница 13: ...xi 920011 649 Operator Manual Table of Contents ...
Страница 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Страница 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Страница 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Страница 167: ......
Страница 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...